|
To ensure that all the profiles in our public access database are
comparable (see navigate
repository), we have spent considerable time
and effort identifying sources of variability in microarray data
(Bakay et al. 2002a), and developing quality control parameters (QC)
and standard operating procedures (SOP). SOPs developed within the
center employ the QC checkpoints at various stages in the typical
microarray workflow from total RNA extraction to GeneChip®
hybridization and scanning. This standardized methodology for quality
assessment has been shown to result in the generation of reliable and
reproducible data. A summary of the threshold values for each QC
criteria follows, with a more detailed explanation of each of the
measures in the next section.
Summary of QC criteria:
- RNA: clearly visible 18S and 28S RNA bands, and A260/280 ratio of
>1.8.
- Input total RNA:
- Standard (one round): 5-10 micrograms for one round
amplifications
- Two-round: 100-200 ng of total RNA input into first round, 100
ng converted cRNA/cDNA into second round.
- IVTs:
- Standard (one round): ouput cRNA = 4-10 fold adjusted
amplification
- Two round: Round 1: 4-10 fold adjusted amplification. Round 2:
400-fold adjusted amplification.
- Biotinylated cRNA integrity: 500-3,000 bp distribution
- Biotinylated cRNA fragmentation: 50-200 bp distribution
- Hybridization (chip) controls:
- Visual inspection of image shows no chip defects
- Percent present calls >30% (A chips)
- Background intensity: <150 (PMT = 1,800)
- Scaling Factor: <5.0 (PMT = 1,800)
- Spike in controls: “present”, BioB<BioC<BioD<CreX
- cDNA integrity: GAPDH 3’/5’ ratio <3.0
The following descriptions summarize our experience with the above parameters.
- RNA integrity and purity: Extracted total RNA is evaluated
on a 1% agarose gel or Agilent Bioanalyzer for presence of 28S and
18S rRNA bands. Ideally, the intensity of the 28S band should be
twice the intensity of the 18S. RNA which is partially degraded will
appear smeared, and lack rRNA bands. RNA concentration is quantitated
by UV spectrophotometry at an absorbance of 260, or using an Agilent
Bioanalyzer. The A260/280 ratio is used as a measure of RNA purity.
RNA with an A260/280 ratio of 1.8-2.1 is considered acceptable for
expression profiling. RNA that does not meet these QC criteria
typically fail the IVT fold-amplification threshold at a high rate
(see below).
- RNA quantity: For single round amplifications
(one IVT reaction), between 3 and 10 micrograms of total RNA is used
(see required fold-amplifications below). For two round IVTs from
small samples, 100-200 ng of total RNA is input into the first round,
and 100 ng of adjusted cRNA input into the second round. Note that a
common reason for failure of two round amplifications is the use of
too much input cRNA into the second round.
- IVT fold-amplification: We have found that the
single most critical QC measure is the efficiency of amplification
from the in vitro transcription reaction. Inadequate amplification
reactions lead to low amounts of biotinylated cRNA, and these samples
typically fail subsequent chip-based QC measures. Inefficient IVT
reactions are also the major cause for rejection of samples, as they
can be variable. We required that each IVT (from either 1 round or 2
round amplifications) give a minimum of a 400-fold amplification of
polyA+ mRNA. For 1 round amplifications, each sample must achieve
four-fold amplification from the starting total RNA amount to the
adjusted cRNA amount (as polyA+ is only ~1% of total RNA, this
translates to a 400-fold amplification of polyA+ mRNA). For example,
a sample started from 6ug of total RNA must have no less than 30ug
after the IVT (30ug of cRNA- 6ug starting material = 24ug LABELED
RNA, i.e. four-fold amplification). Additionally, the Microarray
Center tracks the IVT fold-amplification for all samples within a
project for consistency. Samples with abnormally high or low
amplifications are flagged in subsequent data analysis. For 2 round
amplifications, 100-200 ng of input total RNA should give a minimum
of 400 ng of unlabeled cRNA. When this is converted to
double-stranded cDNA, 100 ng of this cDNA is input into the second
biotinylated IVT, and must provide 40 micrograms of biotinylated
cRNA. Note that the adjusted output of the first round is considered
“pure polyA+ cDNA”, and thus there is a 400-fold amplification
required.
- cRNA integrity: cRNA is evaluated for quality on
an agarose gel or Agilent Bioanalyzer similar to the way that total
RNA was assessed. The cRNA gel should show a smear of RNA ranging
from 500-3000 bases. A large abundance of RNA around 100bp indicates
degradation of the RNA while intensely bright wells indicate
significant genomic DNA contamination.
- cRNA fragmentation: Proper base hydrolysis of
the cRNA is also checked by agarose gel or Agilent Bioanalyzer.
Fragments must range from 50-200bp in size. Smaller fragments
indicate over fragmentation and will result in unreliable
hybridization, larger fragments indicate insufficient fragmentation.
- Hybridization efficiency:
GeneChip® hybridization is assessed by several parameters, namely: %
present calls, background intensity, scaling factor and spike-in
controls.
- Visual inspection of array hybridization: Microarray images are
manually inspected for any chip defects (scratches, holes). Chips
with obvious defects are discarded, and the hybridization solution
run on a new chip.
- Percent present calls: As per Affymetrix suggestion, the
Microarray Center requires that all human, murine and rat A chips
have a minimum of 30% of the represented genes called “present”. B
series chips are required to have a minimum of 20% present calls.
Note that older U95 series arrays (B, C, D, E) have less present
calls due to ambiguous ESTs and poorly performing probe sets (see
Bakay et al. 2002b )
- Background intensity: On GeneChip® scanners with PMT values set
at 1800, the Microarray center requires that all chips have a
background intensity of no more than 150. On GeneChip® scanners with
PMT values set at 18,000 the Microarray Center requires that all
arrays have a background intensity of no more than 1500.
Additionally, the Microarray Center tracks the background
intensities of all GeneChips® within a project and flags samples
with background intensities outside the range of others in the
project.
- Scaling factor: Also as per Affymetrix suggestion, the
Microarray Center requires that all GeneChips® used for analysis
have a scaling factor of no more than 5.0 when GeneChip® scanners
are tuned at high (18000) or low (1800) PMT voltages with a target
intensity respectively of 800 and 150.
- Spike-in controls: As described in the Affymetrix protocol, all
GeneChip® hybridizations include the use of spike-in controls,
namely, BioB, BioC, BioD, and CreX. The Microarray Center requires
that all chips used in analysis have ‘present’ calls for BioC, BioD,
and CreX. Additionally, hybridization efficiency is only considered
acceptable when these genes are detected in increasing quantities:
BioB<BioC<BioD<CreX.
- cDNA integrity: The quality of hybridized material is assessed by
use 3’ and 5’ ends endogenously expressed housekeeping controls.
Specifically the center routinely uses GAPDH for human, murine, and
rat chips. Only GeneChips® with ratios <3.0 are considered analysis.
|